Industrial Conductivity Meter High Accuracy

Modello numero.
DDG-3080

Stability of The Electronic Unit
±0.2%Fs±1 Unit/24h

Output Current Error
: ≤±l%Fs

Error of Electronic Unit Caused by Ambie
≤±0.5%Fs

Ambient Temperature
0~60ºC

Pacchetto di trasporto
Carton Box

Specifica
146 x 146 x 150 mm

Marchio
BOQU

Origine
Cina(Terraferma)

Codice HS
9027809900

Capacità produttiva
300 Sets/Week

Descrizione

DDG-3080 Industrial Conductivity Meter

DDG-3080 Industrial Conductivity Meter is our company’s latest top-class instrument equipped with a microprocessor and
complete
English menu. It has display and menu operation pattern; it is characterized by its intelligent operation, multi-function
and high measuring performance and its strong environmental adaptability. It can be widely used for continuous monitoring of
the conductivities of solutions used in the following fields: thermal power plant, chemical fertilizer, chemical industry, metallurgy,
environmental protection, pharmacy manufacturing, biochemistry, foodstuff and running water.

Caratteristiche

It has a complete English display and friendly interface. Various parameters can be displayed at the same
time: conductivity, output current, temperature, time and status. Bitmap type liquid crystal display module
with high resolution is adopted. All the data, status and operation prompts are displayed in English. There
is no symbol or code that is defined by the manufacturer.

Technical Indexes

1.Conductivity measuring range: 0.01~20μS/cm (Electrode: K=0.01)
0.1~200μS/cm (Electrode: K=0.1)
1.0~2000μS/cm (Electrode: K=1.0)
10~20000μS/cm (Electrode: K=10.0)
30~600.0mS/cm (Electrode: K=30.0)
2. Intrinsic error of the electronic unit: conductivity: ±0.5%FS, temperature: ±0.3ºC
3. Range of automatic temperature compensation: 0~199.9ºC, with 25ºC as the reference temperature
4. Water sample tested: 0~199.9ºC, 0.6MPa
5. Intrinsic error of the instrument: conductivity: ±1.0%FS, temperature: ±0.5ºC
6. Automatic temperature compensation error of the electronic unit: ±0.5%FS
7. Repeatability error of the electronic unit: ±0.2%FS±1 Unit
8. Stability of the electronic unit: ±0.2%FS±1 unit/24h
9. Isolated current output: 0~10mA ( load<1.5kΩ)
4~ 20Ma (load<750Ω) (double-current output for option)
10. Output current error: ≤±l%FS
11. Error of electronic unit caused by ambient temperature: ≤±0.5%FS
12. Error of the electronic unit caused by the supply voltage: ≤±0.3%FS
13. Alarm relay: AC 220V, 3UN
14. Communication interface: RS485 or 232 (opzionale)
15. Alimentazione elettrica: AC 220V±22V, 50Hz±1Hz, 24VCC (opzionale)
16. Protection grade: IP54 , aluminium shell suitable for outdoor use
17. Clock accuracy: ±1 minute/month
18. Data storage capacity: l month (1 point/5 minutes)
19. Saving time of data under a continuous power-failure condition: 10 anni
20. Overall dimension: 146 (lunghezza) x 146 (width) x 150 (depth) mm; dimension of the hole: 138 x 138mm
21. Working conditions: ambient temperature: 0~60ºC; relative humidity <85%
22. Peso: 1.5kg
23. The conductivity electrodes with the following five constants are usable: K=0.01, 0.1, 1.0, 10.0, E 30.0.

Industrial Conductivity Meter High AccuracyIndustrial Conductivity Meter High AccuracyIndustrial Conductivity Meter High AccuracyIndustrial Conductivity Meter High AccuracyIndustrial Conductivity Meter High AccuracyIndustrial Conductivity Meter High AccuracyIndustrial Conductivity Meter High Accuracy