Beschreibung
This product is suitable for static testing of various types of coils/stators, widely used in the testing of coils and stators of electric vehicles, shaded pole motors, universal motors, asynchronous motors, tubular motors, linear motors, and automated production lines.
Compatible: six-in-one test items such as withstand voltage, insulation, resistance, and turn-to-turn.
High performance: voltage regulation, insulation, turn-to-turn and other voltage adjustments using closed-loop voltage feedback adjustment system to accurately control the voltage regulation and rising process, stable voltage output.
Module: Module test unit to avoid interference, and customize single/multi-station, serial/parallel test system as required.
Intelligent: intelligent self-check, remote diagnosis and online upgrade, quick location of problems and eliminating.
Easy maintenance: plug-in design, convenient for disassembly; intelligent self-check, remote diagnosis, online upgrade and other flexible settings, quick location of problems.

Tested items
Stator: withstand voltage, insulation, resistance, resistance balance, interturn, inductance, rotation

| Application industries | Various types of coils and stators | ||
| Model of product | AN8310S(F) | Custom parameters | |
| Withstand voltage test | Output voltage | Range: 300 ~ 3000 VAC, step: 10v | 5000 VAC |
| Allowable error: ± (2.5% x setting +10V) | |||
| Break down current | Range: 0.10~20.00 mA | 100 mA | |
| Allowable error: ± (2.5% x reading +5 words) | |||
| Dauer | Range: 1.0~99.9s, allowable error: ± (1% x setting +0.1s) | ||
| Insulation test | Output voltage | Range: 200 ~ 1000 VDC, step: 5v | |
| Allowable error: ±(3% x setting +5V) | |||
| Ripple coefficient | <5% | ||
| Resistance measurement | Range: 1.0~500MΩ | ||
| Allowable error: ≤200MΩ, ±(3% x reading +5 words) | |||
| >200MΩ, ±(5% x reading +8 words) | |||
| Dauer | Range: 1.0~99.9s, allowable error: ± (1% x setting +0.1s) | ||
| Resistance test | Test and measurement | Range: 0.1~20k Ω, unit: Ω | 1mΩ~400KΩ |
| Allowable error: ±(0.3% x reading +3 words) | |||
| Temperature measurement | (0.0~60.0)°C, allowable error: ±0.5ºC | ||
| Resistance balance | Calculation formula | |Resistance -Resistance average|/Resistance average*100% | |
| Display range | 0.0%~99.9% | ||
| Interturn test | Pulse voltage | Range: 300~3000 V, step: 100v | 5000 v |
| Allowable error: ±(3% x setting +10V) | |||
| Wavefront time | ≤0.5µs | ||
| Acquisition frequency | 5kHz~40MHz | ||
| Wave parameters | Area, difference area, (1~10) consecutive impacts; range: 0~99% | ||
| Inductance test | Test range | 0.0001μH-99.99kH | |
| Allowable error | 0.5% | ||
| Test frequency | 100Hz,120Hz,1kHz,10kHz | ||
| Test level | 0.1,0.3,1.0(v) | ||
| Rotation test | Decision parameter | Forward rotation, reverse rotation, no rotation | |